University College London
Browse

Optical ultrasound image correction with fibre-optic shape sensors - data

Download (60.86 MB)

Significance: Optical ultrasound (OpUS) imaging, in which ultrasound is generated and received optically, is well suited to guiding minimally procedures. 2D images can be synthesized by translation of an OpUS probe within the body. If the probe positions are not known precisely, spatial distortions in the images may occur. Fibre-optic shape sensors offer the possibility to reduce distortions by providing real-time probe positions in an all-optical, miniaturized device.

Aim: We aimed to correct 2D OpUS images acquired with a forward-looking OpUS probe during scanning at non-uniform speeds, using tip position data from a fibre-optic shape sensor.

Approach: An OpUS probe was combined with a fibre-optic shape sensor to allow for real-time acquisition of both ultrasound depth scans and probe tip positions, as the probe was scanned at varying speed across a wire grille imaging phantom. The resulting distorted OpUS images were corrected using the shape sensor position data.

Results: The corrected images corresponded well to the true positions of the wires in the phantom, and distortions in the raw images caused by pauses and varying speeds of the probe movement were largely removed.

Conclusions: Fibre-optic shape sensing is a promising modality for reducing distortions during 2D OpUS images acquired by translating single-element optical transducers.

Funding

Wellcome Trust Centre for Surgical and Interventional Sciences

Wellcome Trust

Find out more...

Nanocomposite materials for sensing in next-generation minimally-invasive medical devices

Engineering and Physical Sciences Research Council

Find out more...

RF/201819/18/125

Wellcome EPSRC Centre for Surgical and Interventional Sciences

Engineering and Physical Sciences Research Council

Find out more...

Ultra-low-cost endoscopy for gastric cancer screening in rural China

Department for Business, Energy and Industrial Strategy

Find out more...

Multispectral polarization-resolved endoscopy and vision for intraoperative imaging of tissue microstructure and function

Engineering and Physical Sciences Research Council

Find out more...

Robotic Assisted Imaging

Engineering and Physical Sciences Research Council

Find out more...

CiET1819/2/36

History